The only apera ph electrodes that must be always stored in storage solutions and can be damaged if stored dry are the ones that use polymer electrolytes such as our probes for ph60s ph60s z labsen 763 and labsen 753 these probes are made with a different type of material and over time will be damaged if stored dry.
Ph electrode storage solution recipe.
Do not store electrode in distilled or deionized water this will cause ions to leach out of the glass bulb and render your electrode useless.
Always keep your ph electrode moist.
To reuse the storage bottle included with the electrode slide the cap and then the o ring onto the electrode insert the electrode midway into the bottle containing storage solution or a 50 50 mixture of 4 m potassium chloride and ph 4 standard buffer and gently screw on cap.
2 while the probe is soaking prepare a batch of new storage solution.
1 let the ph electrode soak for 4 8 hours in an hcl solution between 0 1 and 1 0 m.
The solution you have should be fine if you can change the internal solution on a monthly basis more often if you use it continuously.
A dry probe needs to be hydrated by placing in storage solution before calibration.
The electrode is already filled with 4m kcl saturated with agcl gamry 955 00004.
10 grams of solid kcl dissolved in 100 ml of ph 4 buffer.
A probe that has a hydrated layer will generate a different voltage in a ph buffer than a ph electrode without a hydrated layer.
Most ph electrode storage solutions are 4 m kcl with a ph about 4.
So calibrating a dry electrode will cause inaccurate readings when taking measurements as the hydration layer is formed.
10 grams of solid kcl.
1 let the ph electrode soak for 4 8 hours in an hcl solution between 0 1 and 0 5 m.
The internal solution is also saturated with silver chloride.
Any ph 4 buffer mixture works well but use only kcl for the salt other salts do not work.
If 4 m kcl is not available use a ph 4 or 7 buffer solution.
This solution is appropriate for any of pasco s sensors wit.
I want to prepare an electrode filling and storage solution to use in my ag agcl reference electrode.